6. Set up the network
unuZ@zr
to output the fundamental frequency:
/
.:
...:....;.
.,.
..
. .
a. Press
~$Bj$gi$.
b. Enter the fundamental frequency. For example, press
lso)
m
to enter
the hrst fundamental frequency in the 3
GHz
test record.
7. Measure and record the power in the second or third harmonic by taking a
single sweep with the spectrum
aruzlgzer:
a.
Press @iTZVKQ.
b. Read the MARKER A measurement, and record it in the appropriate row
of the test record under
Measurement Value
(dBc).
8. Reset the
spectrum
un&z@r marker:
9. Repeat steps 5 through 8 for the remaining second and third harmonic
frequencies, and the fundamental frequencies listed in the test record.
System Verification and
2-111
Performance Tests