Post-Repair Procedures for HP 8753E
lhble 3-l lists the additional service procedures which you must perform to
ensure that the instrument is working correctly, following the replacement of an
assembly. These procedures can be located in either Chapter 2 or Chapter 3.
mform
the procedures in the order that they are
listed
in the table.
‘able
3-l.
Related Service Procedures
Replaced
Assembly
Al Front Panel
Keyboard
A2
Front Panel
[nterface
A@stments/
VerUlcation
Correction Constants (Ch.
3)
(a.
2)
None ServiCl?flestO
Service
Test
23
None
Service
lbt
0
Service
l&t
23
service
Test
12
‘Rst.s
66
-
80
k3
Source
A9
Switch Positions
lbt
Port Output Frequency Range
Source Def CC
(!&St
44)
and Accuracy
Pretune Default CC
(!&St
45) Test Port Output Power Accuracy
AnalogBusCC(Test46)
‘l&t
Port Output Power Range and
Source Pretune CC
(Test
48)
Linearity
RF Output Power CC (‘l&t 47)
Tl?st
Port output/Input
Hamlonics
Sampler
Maguitude
and Phase
CC
(‘l&t 53)
(Option
002
only)
Cavity OscUator Frequency CC
(T&t
54)
Source Spur Avoidance Tracking
EEPROM Backup Disk
WAS/A6
Samplers
A9
Switch Positions
MInimmu
R Channel Level
Sampler Maguitude and Phase CC
(l&t
53)
(if R sampler replaced)
IF Amplifier CC
(‘l&t
51)
%stPortCroWalk
EEPROM Backup Disk
Test
Port Input Frequency Response
A7
Pulse Generator
A9
Switch Positions
Test
Port Input Frequency Response
Sampler Magnitude and Phase CC
(Test
53) Test Port Frequency Range and
EEPROM
lb&up
Disk
A-
A8
Post Regulator
A9
Switch Positions
Service
l&t
0
Cavity
Oscillabr
Frequency CC
(l&t
54)
Check
A8 test point voltages
Source Spur Avoidance Tracking
EEPROM Backup Disk
3-2
Adjustments and Correction Constants