A9
Switch Positions
1. Remove the power line cord from the analyzer.
2. Set the analyzer on its side.
3. Remove the two lower-rear comer bumpers from the bottom of the
instrument with the
T-10
TORX screwdriver.
4. Loosen the captive screw on the bottom cover’s back edge, using a
T-15
TORX screwdriver.
5. Slide the cover toward the rear of the instrument.
6. Move the switch as shown in
Figure
3-l:
w
Move the
A9
switch to the Alter position before you run any of the
correction constant adjustment routines. This is the position for altering
the analyzer’s correction constants.
w
Move the
A9
switch to the Normal position, after you have run correction
constant adjustment routines. This is the position for normal operating
conditions.
7.
Reins&R
the bottom cover, but not the rear bumpers
Adjustments and Correstion Constants
3-5