Analog Bus Correction Constants (‘I&t 46)
Aruzlgmr
warmup
time:
30 minutes.
This procedure calibrates the analog bus by using three reference voltages
(ground, + 0.37 and +2.5 volts), then stores the calibration data as correction
constants in
EEPROMs.
2. Observe the analyzer for the results of the adjustment routine:
n
If the analyzer displays
ABUS
Cor DONE, you have completed this
procedure.
n If the analyzer displays
ABUS
Cor FAIL, refer to Chapter 6, “Digital Control
Troubleshooting.
n
Adjustments and Correction Constants
3-8