Source Pretune Correction Constants
(Tkst
48)
An.u&er
warmup time: 30 minutes.
This procedure generates pretune values for correct phase-locked loop
operation.
2. Observe the analyzer for the results of this adjustment routine:
w
If the analyzer displays
Pretune
Cor DONE, you have completed this
procedure.
n If the analyzer displays FAIL, refer to Chapter 7, “Source
Troubleshooting.
n
3-l
0 Adjustments and Correction Constants