IF Amplifier Correction Constants
(‘I&t
61)
Required Equipment and
‘lbols
Antistatic Wrist Strap
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 9300-1367
Antistatic Wrist Strap Cord
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 9300-0980
Static-control
Table
Mat and Earth Ground Wire . . . . . . . . . . . . HP P/N
9300-0797
Additional Required Equipment for 50 ohm Analyzers
RF Cable
-
(5OQ)
24-inch,
APC-7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 8120-4779
Additional Required Equipment for
75Q
Analyzers
RF Cable
-
(750)
24-inch,
Type-N
. . . . . . . . . . . . . . . . . . . . . . . . . . . . HP P/N 8120-2408
Anulgz~warrnupTime:
30 minutes.
These correction constants compensate for possible discontinuities of signal
greater than -30
dBm.
1. Connect the RF cable from Port 1 to Port 2 of the analyzer.
3. Observe the analyzer for the results of the adjustment routine:
z If DONE is displayed, you have completed this procedure.
w
If FAIL is displayed, check that the RF cable is
COMeCted
from Port 1 to
Port 2. Then repeat this adjustment routine.
w
If the analyzer continues to fail the adjustment routine, refer to the
“Digital Control Troubleshooting” chapter.
3-l
6 Adjustments and Correction Constants

Table of Contents