EasyManua.ls Logo

HP 8753E - Page 203

HP 8753E
612 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25. Enter the insertion loss of the through cable (determined in step 15) and
press
:~~~~~~.
For example, if the insertion loss of the through cable at
.....
. . . . .
....
i
..:::.~.....~........~...~i...~;;......~..
1
GHz
is found to be 0.25
dB,
then press 1.25)
@.
26. When the analyzer completes the test, observe the display for the results:
n If you see Sampler Cor
-
DONE, you have completed this procedure.
n
Ifyousee Sampler Cor
-
FAIL, it is necessary to adjust the sampler gain
offset values, which are stored in EEPROM.
b.
Enter the new value at the accessed address by pressing
$jpKK:
(46)
a).
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.y:.
_
.,.,
‘..~..~.~.~,~,.,.
. . . . . . . ..
...:::::..
C-
Access the second address by pressing
~~~~~~~-:~~~~
c-1
Ixl).
d.
Enter the new value at the accessed address by pressing
$&%I,@
@
@.
e. Press
(FEZ]
for the analyzer to use the new values.
f. Repeat the “Sampler Correction Constants Routine” starting at step 16.
n If the analyzer continues to fail this adjustment routine, refer to Chapter
7, “Source Troubleshooting.
n
d.
Enter
the
new
value
at the accessed address by pressing
@$&I$:
(248_)
jxl.
e. Press
B
for the analyzer to use the new values
3-26 Adjustments and Correction Constants

Table of Contents

Other manuals for HP 8753E

Related product manuals