observe the A10 analog bus nodes, perform step AlO, below. Then follow the
node-specific instructions
Step
AlO.
Press:
Node9
+0.37
V(+O.37
V reference)
.~,,‘....:..~:...:.:.:....~I.~:~:.:.’
.
.
.
.
.
::.::.:.:::.:
;..:
i~.~.~.~.~.~.~
~:::::
Perform step
AlO,
above, and then press
IMeas)
~~~~~~~~~:
;
Check for a flat line at approximately + 0.37 V. This is used as the voltage
reference in the “Analog Bus Correction Constants” adjustment procedure. The
voltage level should be the same in high and low resolution; the absolute level is
not critical.
Node10
+2.50
V(+2.50
V reference)
Service
Key
Menus and Error Messages
lo-33