PWR MTR: NOT ON/CONNECTED OR WRONG ADDRS
Error Number The power meter cannot be accessed by the analyzer. Verify
117
that the power meter address and model number set in the
analyzer match the address and model number of the actual
power meter.
SAVE FAILED. INSUFFICIENT MEMORY
Error Number You cannot store an instrument state in an internal register due
151
to insufficient memory. Increase the available memory by
clearing one or more save/recall registers and pressing
m,
or
by storing
files
to a disk.
SELFTFST
#n
FAILED
Service Error Internal test #n has failed. Several internal test routines are
Number 112 executed at instrument preset. The analyzer reports the first
failme detected. Refer to the internal tests and the
self-diagnose feature descriptions earlier in this chapter.
SOURCEPOWERTLJRNElDOFF,
RESET UNDER POWERMENU
Information
You have exceeded the maximum power level at one of the
Message
inputs and power has been automatically reduced. The
annotation
P+
indicates that power trip has been activated.
When this occurs, reset the power and then press
LMenu)
&J@
. . . . . . . . .
;.;,.,.,.,;
. . . . . . .
.::
..
. . . . . .
. . . . . . . . . . . . . . . . .
.::::.
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
,,.........
.;...
.~~~~~:.~~~~~~~~~~.
t
to
sdtc.,
on
the
power.
This
message
follows error numbers 57, 58, and 59.
Service
Key
Menus and Error Messages
lo-59