4.
Test
Port Output Power Range and Linearity
Specifications
Power Range
1
Power Level Linearity1
1
1
-15 to +5
dBm
I
f0.2
dB
1+5to
+10dBm21
f0.5
dB
I
+5
to +8
dBm3
I
f0.5
dB
1
Relative to 0
dBm
output
level.
2 Applies
to
instruments not using Option 076.
3
For Option
075 only.
Required Equipment
Power Meter
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
8482A
Adapter,
APC-7
to Type-N
(f)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
11524A
Additional Required Equipment for Analyzers with Option 006
Power Sensor
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
8481A
Additional Required Equipment for Analyzers with Option 075
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
8483A
Option
HO3
Analgzm warmup time:
1
hour
Perform this test to verify the analyzer’s test port output power range and
power level linearity at selected CW frequencies.
System Verification and
2-27
Performance Tests

Table of Contents