1. Zero and calibrate the power meter. Refer to the power meter operating
manual for more information on how to do this task.
2. On the network analyzer, press
(GJ
LG_)
~~.~~~~~~
Isdo_)
Lk/m.
Set the
power meter calibration factor for this CW frequency.
3. Connect the equipment as shown in
Figure
2-11.
PORT 2
P
ADAPTER
APC-7
to N(F)
HP
8481A
POWER SENSOR
3
-
6GHz
HP
8482A
POWER SENSOR
30KHz
-
3GHz
HP
0753E
NETWORK ANALYZER
HP
438A
HP
8753E
NETWORK ANALYZER
WITH OPTION 075
POWER SENSOR
HP
438A
*
DIRECT CONNECTION
/
sg61
le
Figure 2-11.
‘l&t
Port Output Power Range and Accuracy
‘Ikst
Setup
2-28
System Verification and
Performance Tests