2. Using an ohmmeter, verify that the RF cable is not open. In addition,
examine both the cable connectors-measure the resistance between the
cable center pin and the cable connector and make sure it is
not
close to
zero.
3. Check the R sampler by substituting it with the B sampler
(A6).
a. Move cable
W8
to the B sampler
(A6),
as shown in
Figure
2-14.
All
sg6115-z
Figure 2-14. Connections for Substituting the
B
!kmpler
(A4)
4. Connect the equipment as shown in
F’igure
2-15.
System Verification and
2-35
Performance Tests