Figure 2-15. Setup for Checking the B Sampler
(A4)
5. Repeat the test, but select the B sampler
(A6)
by pressing
m
~:~~~~~~~~:
‘B
h
step
2.
Use
the
following
specific&ions:
..__
._
.
.._
_
_
_i
.,.,,_
_
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
_
300
kHz
to 3
GHz
c-27
dBm
3
GHz
to 6
GHz
c-22
dBm
6. If the analyzer fails the test, replace the All assembly.
7. Verify that the
high/Iow
band adjustments are still within specifications For
more information on how to perform this task, refer to the “High/Low Band
Transition Adjustment” located in Chapter 3, “Adjustments and Correction
Constants
n
8. Refer to Chapter 7, “Source Troubleshooting,” for more troubleshooting
information.
2-36 System Verification and
Performance Tests