EasyManua.ls Logo

HP 8753E - Page 73

HP 8753E
612 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.
‘Jkst
Port Input Noise Floor Level
Specifkations
Frequency
Range
‘l&t
Port
IF
Bandwidth
Average
Noise Level
300
kHz
to 3.0
GHz
Port
1
3kHz
-82
dBm
300
kHz
to 3.0
GHz
Port
1
10 Hz
-102
dBm
300
kHz
to 3.0
GHz
Port
2
3
kHz
-82
dBm
300
kHz
to 3.0
GHz
Port
2
10 Hz
-102
dBm
3.0
GHz
to 6.0
GHz’
Port
1
3kHz
-77
dBm
3.0
GHz
to 6.0
GHzl
Port
1
10 Hz
-97
dBm
3.0
GHz
to 6.0
GHzl
Port
2
3kHz
-77
dBm
3.0
GHz
to 6.0
GHz’
Port
2
10 Hz
-97
dBm
1
Only for analyzer with Option 006
-
30
kHz
to 6
GHz
range.
Equipment Required for
508
Analyzers
Calibration Kit,
7-mm
. . . . . . . . . . . . . . . . . . . . . . . . . , , . . . . . . . . . . . . . . . . . . . . . HP 85031B
Equipment Required for 75 ohm Analyzers
Calibration Kit, Type-N
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP
85036B
An&zerwarmuptime:1
hour
Perform this test to determine the HP 8753E port 1 and port 2 noise floor levels
at the input test ports
Syctem Verification and