Keysight N9038A MXE Service Guide 101
Instrument Messages
Event Messages
–241 Hardware missing;
Option not installed
The optional hardware is not installed
–250 Mass storage error; A problem was found with the mass storage device (memory, disk drive, etc.).
The exact problem cannot be specifically identified
–250 Mass storage error;
Access denied
Access is denied
–250 Mass storage error;
Bad path name
The specified path is invalid
–250 Mass storage error;
Can only import
single trace .csv
files
Trace files containing multiple traces can not be imported. However, if you
need to recall multiple traces you can use the
Save and Recall functions,
rather than the Import and Export functions
–250 Mass storage error;
Can only load an
Antenna Unit into
Correction 1
The only Correction register that supports Antenna Units is number 1. You have
attempted to load an Ampcor file that contains antenna units into another
register
–250 Mass storage error;
Cannot make
The directory or file cannot be created
–250 Mass storage error;
Different Antenna
Unit already in use
Attempt to import Corrections file with Antenna Unit that differs from an in-use
correction
–250 Mass storage error;
Directory not found
The system cannot find the path specified
–250 Mass storage error;
Failed to Load trace.
Bad file format.
The load trace operation could not be completed, as the input file was not in
the expected format. You can only load traces that were previously saved using
the
Save, Trace feature
–250 Mass storage error;
File <filename> wrong
type
Attempt to import a data file that is not the proper type for this operation
–250 Mass storage error;
File <filename> and
instrument version
mismatch
While opening a file, there was a mismatch between file version or model
number with instrument version or model number. The import still tried to load
as much as possible, but you should check it closely
–250 Mass storage error;
File contains
incorrect data for
this operation
There is a mismatch between the file data type of the file specified and the
destination indicated. For example, a correction set cannot be loaded/imported
into a limit line
–250 Mass storage error;
File empty
Cannot save trace because it contains no data. Check that the trace is turned
on and contains some valid data
Err# Message Verbose/Correction Information