102 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–250 Mass storage error;
Invalid register
number for *SAV or
*RCL Mass Storage
error
You have used the *SAV command to save a state to a non existent state
register.
Or
You have used the *RCL command to recall a state register that wasn’t
previously saved with the *SAV command
–250 Mass storage error;
Lock violation
The process cannot access the file because another process has locked a
portion of the file
–250 Mass storage error;
Mkr Table must be on
to save Mkr Table as
Meas Results
You have to have a Marker Table on the screen before you can save it. Turn on
the Marker Table and try again
–250 Mass storage error;
No file names
available
Attempt to use the auto file name generation when all 10,000 file names are
taken
–250 Mass storage error;
Open failed
The system cannot open the device or file specified. This could be because the
storage media is full, or possibly due to a filename error. If using an external
storage device, check that the device is properly formatted
–250 Mass storage error;
Pk Table must be on
to save Pk Table as
Meas Results
You must have a Peak Table on the screen before you can save it. Turn on the
Peak Table and try again
–250 Mass storage error;
Read fault
The system cannot read from the specified device
–250 Mass storage error;
Register <number>
empty
Attempt to recall a register with nothing in it
–250 Mass storage error;
Sharing violation
The process cannot access the file because it is being used by another process
–250 Mass storage error;
Spectrogram must be
on to save as Meas
Results
You must have a Spectrogram on the screen before you can save it. Turn on the
Spectrogram and try again
–250 Mass storage error;
Too many open files
The system cannot open the file
–250 Mass storage error;
Write fault
The system cannot write to the specified device
–252 Missing media A legal command or query could not be executed because of missing media
–253 Corrupt media A removable media was found to be bad or incorrectly formatted. Any existing
data on the media may have been lost
–254 Media full A legal command/query could not be executed because the media was full
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals