Keysight N9038A MXE Service Guide            147
Input Selection & Level Control
Input Selection & Level Control Section Troubleshooting
12.Turn on the Marker Delta function of the spectrum analyzer being used.
13.Change the A10 Attenuator B setting to 10 dB by pressing AMPTD, 
Attenuation, Mech Atten, 10 dB.
14.Verify that the signal level on the spectrum analyzer drops by 10 dB (±0.25 
dB) from where it was, as shown in Figure 4-14.
If the signal level is incorrect the most likely cause of the problem is the 
A10 Attenuator B assembly itself. However, before replacing an A10 
Attenuator B assembly for any failure, refer to Chapter 8, “Front End 
Control.” for information on how to verify the control signals to it.
Figure 4-14  A10 Attenuator B −10 dB