EasyManuals Logo

Keysight N9038A Service Guide

Keysight N9038A
745 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #255 background imageLoading...
Page #255 background image
Keysight N9038A MXE Service Guide 255
RF Downconverter Section
RF Downconverter Section Troubleshooting
Figure 6-18 A13 RF Front End Assembly Connections - Option 544
4. Connect a spectrum analyzer to the A13 RF Front End assembly Output
(J7).
5. Tune the spectrum analyzer to 322.5 MHz with a span of 10 MHz.
6. Turn the instrument on and allow it to complete its boot up process.
7. Turn the instrument self-alignment routine off by pressing System,
Alignments, Auto Align, Off.
8. Tune the instrument to 100 MHz with a span of 0 Hz by pressing FREQ,
100 MHz and SPAN, Zero Span.
9. Referring to Figure 6-17 or Figure 6-18, disconnect W38 from the A13 RF
Front End assembly low band Input (J2).
10.Connect a signal generator to the A13 RF Front End assembly low band
Input (J2).
11.Tune the signal generator to 100 MHz with an amplitude of -30 dBm.
12.Verify that the signal amplitude on the spectrum analyzer display is
approximately -25 dBm (5 dB gain), allowing for any additional test cable
loss, as seen in Figure 6-19.

Table of Contents

Other manuals for Keysight N9038A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9038A and is the answer not in the manual?

Keysight N9038A Specifications

General IconGeneral
BrandKeysight
ModelN9038A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals