Assembly Replacement Procedures
RF Area - Option 544
Replacement 
1. Refer to Figure 18-25. If the attenuator bracket was removed, position it 
back into the chassis and replace the four screws (4) to attach it to the 
chassis.Torque to 9 inch-pounds.
2. Position the attenuator in the bracket so that the ribbon connector end is 
toward the front of the instrument.
3. Replace the two screws (3) (0515-0372) that attach the attenuator to the 
bracket. Torque to 9 inch-pounds.
4. Replace the ribbon cable W3 and semi-rigid cables W25, W28, W29, and 
W30 to the attenuator. Torque the semi-rigid cables to 10 inch-pounds.
5. Refer to Figure 18-24. Position the Chassis Right Side Outer bracket onto 
the chassis and replace the eleven screws (0515-0372). Torque to 
9 inch-pounds.
6. Replace the top brace. Refer to the Top Brace and Power Supply Bracket 
replacement procedure.
7. Replace the front panel. Refer to the Front Frame Assembly replacement 
procedure.
8. Replace the instrument outer case. Refer to the Instrument Outer Case 
replacement procedure.