9/12/00 B-25
Brackets [ ] indicate optional characters or operands.
<cr> = carriage return; <lf> = line feed; _ = space
Querying Settings
The current setting is retrieved with the Query (Q) com-
mand. The syntax for the query command is
Q[uery ]setting_number[,options] where setting_number is
the number of the desired setting and options is a number to
select the appearance of the response.
Brackets [ ] indicate optional characters or operands. <cr> =
carriage return; <lf> = line feed; _ = space
Option flags elicit the following responses:
Flag values may be added together for desired combinations:
Syntax Response
Q[uery]item_number[,flag] Depends on flag
Flag Response
none Current setting
1 Setting name and current setting
2 Current setting (including spaces) in brackets or parentheses
3 Setting name and current setting (including spaces) in brackets or parentheses
32 Option number for option setting
Example: Response
1.Q74 No
2.Q74, 1 Excd History Enable=No
3.Q74, 2 [_No]

Table of Contents

Other manuals for Larson Davis System 824

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Larson Davis System 824 and is the answer not in the manual?

Larson Davis System 824 Specifications

General IconGeneral
BrandLarson Davis
ModelSystem 824
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals