9/12/00 B-65
Monitor Number
The monitor number is used to specifically address the mon-
itor.
824
Phone Dialing Procedure
The Model 824 dialing process is as follows:
Step 1
The Model 824 recognizes an exceedance or low
memory condition.
Example:
824 sends: ATDT 1-
555-1234(Enter)
Step 2
The Model 824 asks the modem to dial the phone
number.
Modem sends: 824A0123:001:2
(Enter)
Step 3
The modem informs the Model 824 that a connec-
tion has been made. The Model 824 modifies its
baud rate to that of the connected modem. If the
connection is unsuccessful, the procedure is
retried in 4 min.
Step 4
The Model 824 sends the following announce-
ment: “
824: 824A0123:001:2
” (enter), where
(0123) is the Model 824’s serial number, (001) is
the monitor number entered, and (2) is a flag: 1-
Alarm 2-Exceedance 4-Low Memory. More than
one of these can be set at a time, in which case x is
the sum of all set flags.
, will pause dialing for 2 s
space or - used to make the number more read-
able.
T9W 1-412-555-
1212
indicates to use tone dialing, dial 9
for an outside line, wait until the dial
tone is detected, and then dial the
number.

Table of Contents

Other manuals for Larson Davis System 824

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Larson Davis System 824 and is the answer not in the manual?

Larson Davis System 824 Specifications

General IconGeneral
BrandLarson Davis
ModelSystem 824
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals