11/8/00 Using ID Setting Files 7-37
Long -
The “long” interval report (including L
max
,
L
min
, SEL, L
eq
, Pk-I, Pk-II, overloads, exceedances,
and six Ln Percentiles) will print.
Hist Report - (LOG, SSA)
The
Hist Report
either enables or disables the printing of the
Time history report. The history report contains the L
eq
, and
optionally one selectable level (Lpk-I, Lpk-II, or Lmax)
basic measurement parameter for each time history period.
This setting provides the following two options:
No -
The history report will not print.
Yes -
The history report will print.
Menu Layout of SSA Settings
The SLM&RTA
Edit Settings
menu has 49 configurable
items. Provided in this section is a table depicting the menu
layout of the SLM&RTA instrument’s Settings.
Menu
Items
Submenu Available Settings
Options
Title User customized text
entry
SLM Detector Slow
Fast
Impl
Weighting A
C
FLT
Gain A
C
FLT
Transducer Condnsr
Elctret
Direct