7-30 824 Reference Manual 11/8/00
either until the instrument is stopped or runs out of memory.
You can set this to any time value in the following range:
NOTE: If set to 00:00:00 then the period
is 0.125 seconds for SSA and infinite for
LOG.
00:00:01 to 99:59:59 (i.e. one second to 99 hours,
59 minutes and 59 seconds)
Intv Time Sync - (LOG, SSA) Intv Time Sync
setting either enables or disables the time
interval sync feature. When enabled, this feature causes the
Interval history storage operations to be synchronized with
the real-time of day clock. This synchronization can occur
on a 1, 5, 15, 20, 30, or 60 minute basis. The
Intv Period
set-
ting should also be set accordingly so that the synchroniza-
tion works properly. When you start a measurement, the
initial storage interval will last for only a fraction of the
determined synchronization until the start of the next time
segment; at which time a new, synchronized storage interval
will begin.
This setting provides the following two options:
No -
Disables interval synchronization.
Yes -
Enables interval synchronization.
Intv Save Ln’s - (LOG, SSA) Intv Save Ln’s
either enables or disables the automatic stor-
age of the six L
N
percentile levels in the interval record.
This setting provides the following two options:
No -
The six L
N
levels are not stored with the interval
record.
Yes -
The six L
N
levels are stored with the interval-
record.
NOTE: If Intervals are enabled and Interval Ln’s are enabled and
Ln’s are not enabled then Ln’s are turned on and the warning “Set-
ting Conflict” is displayed.
Intv Auto Stop - (LOG, SSA)
NOTE: Do not use this feature with the
Time Sync feature, or you will not get the
full time period you desire.
Use the
Intv Auto Stop
setting to enable or disable the auto-
stop feature. When enabled, this feature causes the instru-
ment to automatically stop the measurement at the end of
each storage interval. This allows you to precisely time a