EasyManua.ls Logo

Larson Davis System 824 - Logic Output #2; Logic Output #2 Time Setting; Heater on; E. A. Cal Tone

Larson Davis System 824
536 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-14 824 Reference Manual 11/8/00
NOTE: If set below 5 seconds, the output
will remain “ON” as long as the condi-
tions are true, and turn “OFF” x sec-
onds (0 to 5) after the condition; if set to
6 seconds or greater, the output will stay
“ON” for at least 5 seconds but will turn
“OFF” as soon as the condition
becomes false or the condition has lasted
the number of set seconds (i.e. 0-5 mini-
mum time, 6-255 maximum time).
Logic Output #2 Logic-Out #2
works the same as Logic Output and has the
same options as shown above.
NOTE: When Logic-Out #2 is not off
then the Heater Control setting has no
effect. It is overriden by Logic-Out #2.
Logic Output #2 Time Setting
The
Logic Output #2
Time sets the number of seconds that
the output will remain turned “ON” following a triggering
event. See
Logic Output Time
setting above.
Heater On
The
Heater Control output
selects the functionality of pin 8
of the control connector. It is an “open drain” output such
that when asserted, the Heater output sinks current to
ground; when unasserted no current flows. It has a maxi-
mum rating of 40 volts.
The “
Heater On
” setting provides the following options:
No
- Is unasserted; which is considered OFF (line
can be used as Logic output #2)
Yes
- Is asserted; which is considered ON
E. A. Cal Tone
The
E. A. Cal Tone
setting is used to manually control the
Electrostatic Actuator (E. A.) of an outdoor microphone sys-
tem. It is a CMOS output that drives from 0 to +5 volts
through a 10,000 ohm current limiting resistor to pin 7 of the
control connector; +5 volts is the asserted state.

Table of Contents