11/8/00 Using ID Setting Files 7-5
“LOG and SSA Time History Settings”
“Logging SLM Interval Settings”
“Logging SLM Exceedance History Settings”
“Logging SLM Daily History Settings
“Logging SLM Define Report Settings”
In the following sections, headings for each setting or dis-
play mode will identify in parentheses the particular System
824 instruments to which they pertain, as shown in the fol-
lowing example.
Example: Title Setting - (
SSA, LOG, ISM
)
Common SLM Settings
The SSA, ISM, and LOG instrument types include standard
SLM functions. This section covers the available SLM set-
tings for these instruments.
Title Setting - (SSA, LOG, ISM)
To change the Title, use the
and
keys to select each charac-
ter. Use the
and
keys to
replace the selected letter. Pressing
the
Reset key clears all charac-
ters from the cursor to the end of the
field.
Use the
Title
setting to create a title or header for printed
reports. This title/header will be placed at the top of each
report you print.
Detector Setting - (SSA, LOG,
ISM)
Detector -
selects the detector rate of the instrument’s RMS
detector. This setting provides the following three options:
The 824 measures all three fre-
quency weightings with three detec-
tors simultaneously and all nine of
these levels can be viewed.
Slow
- An exponential average with a one second
time constant.
Fast
-
An exponential average with a 1
8 second
time constant.