[Zero correction]
The zero point correction in use is displayed.
[Communication error counter]
The communication errors (timeouts) for the load cells are counted here in ascending
order and displayed.
[LC 1…n]
Bar graph display
A
B
C
The bar graph shows three areas:
- A
Dead load (can be changed by calibration)
- B
Maximum capacity E
max
(max. capacity of load cell) including dead load (load cell,
cannot be changed)
- C
Max. usable load including dead load (load cell, cannot be changed)
The colors have the following meanings:
- Red
Weight value is above maximum capacity (overload) or below ¼d
- Green
Weight value is within tolerances
- Orange
Weight value is above maximum capacity E
max
(max. capacity of load cell)
3. Select the desired load cell and press the [Info] soft key.
The load cell data is displayed: