[Info] soft key
Load cell info @admin
Model name
Software version
LC serial number
Emax
n
Y
Z
Overload
Overload counter
Temperature
Max. temperature
Min. temperature
PR6204/50tC3
1.00.04
101
50.0 t
3000 e
14000
3000
75.0 t
0
26.6 °C
36.6 °C
26.4 °C
Display Description
E
max
Maximum capacity
n Max. resolution
Y Minimum LC verication
Z Minimum preload signal recurrence
Overload Weight value above max. usable load
Overload counter Number of weight values above max. usable load.
The higher the number, the higher the probability
of a faulty load cell.
Temperature Current measured temperature
Max. temperature Max. measured temperature
Min. temperature Min. measured temperature
Max. weight value
at
Date and time display
Time of largest load on load cells
Max. weight value Display
Show load cell weight @admin
Zero correction
Communication error count
PR 6204-RU
PR 6204-RO
PR 6204-LO
PR 6204-LU
1.411 t
2.383 t
1.580 t
2.129 t
0.00000 t
40
Info by ID
[by name] soft key
If names are assigned in the menu [System setup]- [Weighing point]- [Calibration]-
[Assign load cell names] , they are displayed.
5 System information Process Controller Maxxis 5 PR 5900