- Function number 5: device type and software version (read), see Chapter 8.2.5.5
- Function number 6: serial number of the weighing point (read), see Chapter 8.2.5.6
- Function numbers 8 to 15: weight data (read), see Chapter 8.2.5.8
- Function numbers 80 to 89: state-controlled action bits (write), see Chapter 8.2.5.9
- Function number 112 to 119; 121: transition-controlled action bits (write), see
Chapter 8.2.5.10
8.2.5.1 Function number 1: scale status (read)
Dynamic status
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 OutOf
Range
Standstill Inside ZSR Center
Zero
Below
Zero
Overload Above Max ADU Error
Byte 1 E9 E6 E1 E3 E7
Byte 2 PowerFail CmdBusy CmdError
Byte 3 TareActive CalActive TestActive
Note:
Byte 0 corresponds to byte 7 in the output area. Weight error in byte 1, see table in
Chapter 9).
Field Function
ADUError AD conversion error (OR function of bits E1, E3, E7).
AboveMax The weight value has exceeded the Max (FSD), but is still within Max
+ permissible overload (gross ≤ FSD+overload).
Overload The weight value has exceeded the measuring range. No valid
weight data is specied (gross > FSD+overload); error 2.
BelowZero The weight value is negative (gross < 0d).
CenterZero The weight value is within center zero (0 ± 0.25 d)
InsideZSR The gross weight value is within the zero setting range.
Standstill The scale is stable.
OutOfRange Below zero or above max. (FSD).
E7 The measuring signal is negative (inverse conversion); error 7
E6 Sense voltage not present or too low; error 6
E3 The measuring signal is >36 mV (no end of conversion); error 3
E1 Arithmetic error (overow); error 1
E9 No communication with xBPI scale; error 9