EasyManua.ls Logo

R&S FSV3000 Series - Page 674

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
674User Manual 1178.8520.02 ─ 01
Bit No. Meaning
5 ALT2 LOWer FAIL
This bit is set if the limit is exceeded in the lower 2nd alternate channel.
6 ALT3 … 11 LOWer/UPPer FAIL
This bit is set if the limit is exceeded in one of the lower or upper alternate channels 3 … 11.
7 CACLR FAIL
This bit is set if the CACLR limit is exceeded in one of the gap channels.
8 GAP ACLR FAIL
This bit is set if the ACLR limit is exceeded in one of the gap channels.
9 to 14 Unused
15 This bit is always 0.
STATus:QUEStionable:EXTended Register
The STATus:QUEStionable:EXTended register contains further status information
not covered by the other status registers of the R&S FSV/A. A separate EXTended
register exists for each active channel.
You can read out the register with STATus:QUEStionable:EXTended:CONDition?
or STATus:QUEStionable:EXTended[:EVENt]?
Table 13-12: Meaning of the bits used in the STATus:QUEStionable:EXTended register
Bit No. Meaning
0 not used
1 INFO
This bit is set if a status message is available for the application.
Which type of message occurred is indicated in the STATus:QUEStionable:EXTended:INFO
Register.
2 to 14 Unused
15 This bit is always 0.
STATus:QUEStionable:EXTended:INFO Register
The STATus:QUEStionable:EXTended:INFO register contains information on the
type of messages that occur during operation of the R&S FSV/A. A separate INFO reg-
ister exists for each active channel.
You can read out the register with STATus:QUEStionable:EXTended:INFO:
CONDition? or STATus:QUEStionable:EXTended:INFO[:EVENt]?. You can
query all messages that occur for a specific channel using the command SYSTem:
ERRor:EXTended? on page 1131.