Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
768User Manual 1178.8520.02 ─ 01
Suffix:
<sb>
.
1..n
sub block number
<ch> 1..n
Tx channel number
Parameters:
<Alpha> Roll-off factor
Range: 0 to 1
*RST: 0.22
Manual operation: See " Weighting Filters " on page 174
[SENSe:]POWer:ACHannel:FILTer[:STATe]:SBLock<sb>:CHANnel<ch>
<State>
This command turns the weighting filter for the specified transmission channel on and
off.
Suffix:
<sb>
.
1..n
sub block number
<ch> 1..n
Tx channel number
Parameters:
<State> ON | OFF | 1 | 0
*RST: W-CDMA: 1, other technologies: 0
Manual operation: See " Weighting Filters " on page 174
[SENSe:]POWer:ACHannel:SBCount <Number>
This command defines the number of sub blocks, i.e. groups of transmission channels
in an MSR signal.
For more information see Chapter 7.3.3.4, "Measurement on Multi-Standard Radio
(MSR) Signals", on page 153.
Parameters:
<Number> Range: 1 to 5
*RST: 1
Manual operation: See " Number of Sub Blocks " on page 169
[SENSe:]POWer:ACHannel:SBLock<sb>:BWIDth[:CHANnel<ch>] <Bandwidth>
[SENSe:]POWer:ACHannel:SBLock<sb>:BANDwidth[:CHANnel<ch>]
<Bandwidth>
This command defines the bandwidth of the specified MSR Tx channel.
This command is for MSR signals only (see CALCulate<n>:MARKer<m>:
FUNCtion:POWer<sb>:PRESet on page 752).
Configuring and Performing Measurements