Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
796User Manual 1178.8520.02 ─ 01
Suffix:
<sb>
.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
ESP:HSP ON
Manual operation: See " Fast SEM " on page 223
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:RESolution <RBW>
This command defines the resolution bandwidth for a SEM range.
In case of high speed measurements, the resolution bandwidth has to be identical for
all ranges.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the measurement range.
Parameters:
<RBW> Resolution bandwidth.
Refer to the data sheet for available resolution bandwidths.
*RST: 30.0 kHz
Default unit: Hz
Manual operation: See " RBW " on page 223
[SENSe:]ESPectrum<sb>:RANGe<ri>:BANDwidth:VIDeo <VBW>
This command defines the video bandwidth for a SEM range.
In case of high speed measurements, the video bandwidth has to be identical for all
ranges.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the measurement range.
Parameters:
<VBW> Video bandwidth.
Refer to the data sheet for available video bandwidths.
*RST: 10.0 MHz
Default unit: Hz
Manual operation: See " VBW " on page 223