Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
801User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STOP <Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the measurement range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STOP -15
For a detailed example see Chapter 14.5.6.11, "Example: SEM
Measurement", on page 823.
Manual operation: See " Abs Limit Start / Stop <n> " on page 225
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt <Level>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit regardless of the limit check
mode.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Relative limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: -50
Default unit: dBc
Example:
SENS:ESP:RANG:LIM:REL:STAR -10
Manual operation: See " Rel Limit Start / Stop <n> " on page 225