Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
802User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:ABS <Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see "Relative limit line functions" on page 214.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the start frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STARt:FUNCtion on page 802).
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
ABSolute -10
For a detailed example see Chapter 14.5.6.11, "Example: SEM
Measurement", on page 823.
Manual operation: See " Rel Limit Start / Stop <n> " on page 225
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:FUNCtion
<Function>
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Function> OFF | MAX
Defines the function to be used to determine the relative limit
line start value
MAX
The maximum of the relative and the absolute level is used as
the limit start value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STARt and [SENSe:
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STARt:ABS commands to define these values.