Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
803User Manual 1178.8520.02 ─ 01
OFF
No function is used, the relative limit line is defined by a fixed
relative start value. Use the [SENSe:]ESPectrum<sb>:
RANGe<ri>:LIMit<li>:RELative:STARtcommand to
define this value.
*RST: OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
FUNCtion MAX
For a detailed example see Chapter 14.5.6.11, "Example: SEM
Measurement", on page 823.
Manual operation: See " Rel Limit Start / Stop <n> " on page 225
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP <Level>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit anytime and regardless of the
limit check mode.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the SEM range.
<li> 1..n
Power class for which the limit is defined.
Parameters:
<Level> Relative limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
For a detailed example see Chapter 14.5.6.11, "Example: SEM
Measurement", on page 823.
Manual operation: See " Rel Limit Start / Stop <n> " on page 225
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:ABS <Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see "Relative limit line functions" on page 214.
Suffix:
<sb>
.
1..n
<ri> 1..n
Selects the SEM range.
Configuring and Performing Measurements