Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
809User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa <Alpha>
This command defines the roll-off factor for the RRC filter.
The RRC filter is available if the power reference is the channel power.
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Alpha> Range: 0 to 1
*RST: 0.22
Manual operation: See " Alpha: " on page 230
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe] <State>
This command turns the RRC filter in the reference range on and off.
The RRC filter is available if the power reference is the channel power.
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Manual operation: See " RRC Filter State " on page 230
[SENSe:]ESPectrum<sb>:RRANge?
This command queries the reference range.
Suffix:
<sb>
.
Sub block in a Multi-SEM measurement
Return values:
<RefRange> Number of the current reference range.
Range: 1 to 30
Usage: Query only
[SENSe:]ESPectrum<sb>:RTYPe <Type>
This command defines the type of the power reference.
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Type> PEAK | CPOWer
Configuring and Performing Measurements