Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
813User Manual 1178.8520.02 ─ 01
You can only use power classes for which limits are defined.
Suffix:
<n>
.
irrelevant
<li> irrelevant
<sb> 1..n
<pc> 1..n
power class
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:LIM:ESP:PCL1 ON
Activates the first defined power class.
Manual operation: See " Used Power Classes: " on page 230
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe] <State>
This command selects the limit check mode for each power class.
Suffix:
<n>
.
irrelevant
<li> irrelevant
<sb> 1..n
<pc> 1..n
power class
Parameters:
<State> ABSolute | RELative | AND | OR
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
Example:
CALC:LIM:ESP:PCL:LIM ABS
Manual operation: See " Used Power Classes: " on page 230