CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum <Level>
This command defines the upper limit of a particular power class.
Note:
The last power class always has an upper limit of 200 dBm.
The upper limit of a power class must always be the same as the lower limit of the
subsequent power class.
The power class must already exist (see CALCulate<n>:LIMit<li>:
ESPectrum<sb>:PCLass<pc>:COUNt on page 812).
Suffix:
<n>
.
irrelevant
<li> irrelevant
<sb> 1..n
<pc> 1..n
power class
Parameters:
<Level> Range: -199.9 dBm to 200 dBm
Default unit: dBm
Example:
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40
dBm.
Manual operation: See " PMin / PMax " on page 231
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum <Level>
This command defines the lower limit of a particular power class.
Note:
The first power class always has a lower limit of -200 dBm.
The lower limit of a power class must always be the same as the upper limit of the
previous power class.
The power class must already exist (see CALCulate<n>:LIMit<li>:
ESPectrum<sb>:PCLass<pc>:COUNt on page 812).
Suffix:
<n>
.
irrelevant
<li> irrelevant
<sb> 1..n
<pc> 1..n
power class
Parameters:
<Level> Range: -200 dBm to 199.9 dBm
Default unit: dBm
Configuring and Performing Measurements