Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
902User Manual 1178.8520.02 ─ 01
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<OptMode> LNOise | LDIStortion
LNOise
Optimized for high sensitivity and low noise levels
LDIStortion
Optimized for low distortion by avoiding intermodulation
*RST: LDIStortion
Example:
INP:ATT:AUTO:MODE LNO
INPut<ip>:EATT <Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (INP:EATT:AUTO OFF, see INPut<ip>:EATT:AUTO on page 902).
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Attenuation> attenuation in dB
Range: see data sheet
Increment: 1 dB
*RST: 0 dB (OFF)
Default unit: DB
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation: See " Using Electronic Attenuation " on page 371
INPut<ip>:EATT:AUTO <State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Setting Basic Measurement Parameters