Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
903User Manual 1178.8520.02 ─ 01
ON | 1
Switches the function on
*RST: 1
Example:
INP:EATT:AUTO OFF
Manual operation: See " Using Electronic Attenuation " on page 371
INPut<ip>:EATT:STATe <State>
This command turns the electronic attenuator on and off.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
INP:EATT:STAT ON
Switches the electronic attenuator into the signal path.
Manual operation: See " Using Electronic Attenuation " on page 371
14.7.3.3 Configuring a Preamplifier
INPut<ip>:EGAin[:STATe]................................................................................................903
INPut<ip>:GAIN:STATe...................................................................................................904
INPut<ip>:GAIN[:VALue].................................................................................................904
INPut<ip>:EGAin[:STATe] <State>
Before this command can be used, the external preamplifier must be connected to the
R&S FSV/A. See the preamplifier's documentation for details.
When activated, the R&S FSV/A automatically compensates the magnitude and phase
characteristics of the external preamplifier in the measurement results.
Note that when an optional external preamplifier is activated, the internal preamplifier is
automatically disabled, and vice versa.
When deactivated, no compensation is performed even if an external preamplifier
remains connected.