Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
923User Manual 1178.8520.02 ─ 01
*RST: AC
Example:
INP:COUP DC
Manual operation: See " Input Coupling " on page 312
INPut<ip>:DPATh <State>
Enables or disables the use of the direct path for frequencies close to 0 Hz.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> AUTO | OFF
AUTO | 1
(Default) the direct path is used automatically for frequencies
close to 0 Hz.
OFF | 0
The analog mixer path is always used.
Example:
INP:DPAT OFF
Manual operation: See " Direct Path " on page 313
INPut<ip>:FILTer:YIG[:STATe] <State>
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1 (0 for I/Q Analyzer, GSM, VSA, Pulse, Amplifier,
Transient Analysis, DOCSIS and MC Group Delay
measurements)
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation: See " YIG-Preselector " on page 313
INPut<ip>:IMPedance <Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50 Ω are supported.
Suffix:
<ip>
.
1 | 2
irrelevant