CALibration:PADJust[:STATe]
Activates or deactivates the preselector adjustment.
This command is only available for instrument modelsR&S FSV/A43/50/67, for fre-
quency sweeps in the Spectrum application.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See "Preselector Adjust" on page 314
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S FSV/A after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the STAT:QUES:POW status register) and the
INPUT OVLD message in the status bar are cleared.
(See STATus:QUEStionable:POWer[:EVENt]? on page 1154 and "STA-
Tus:QUEStionable:POWer Register" on page 677).
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see "RF Input Protection" on page 312.
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:COUPling <CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CouplingType> AC | DC
AC
AC coupling
DC
DC coupling