Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
924User Manual 1178.8520.02 ─ 01
Parameters:
<Impedance> 50 | 75
numeric value
User-defined impedance from 50 Ohm to 100000000 Ohm
(=100 MOhm)
User-defined values are only available for the Spectrum applica-
tion, the I/Q Analyzer, and some optional applications.
*RST: 50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation: See " Impedance " on page 313
See " Unit " on page 370
INPut<ip>:IMPedance:PTYPe <PadType>
Defines the type of matching pad used for impedance conversion for RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<PadType> SRESistor | MLPad
SRESistor
Series-R
MLPad
Minimum Loss Pad
*RST: SRESistor
Example:
INP:IMP 100
INP:IMP:PTYP MLP
Manual operation: See " Impedance " on page 313
INPut<ip>:SELect <Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S FSV/A.
If no additional input options are installed, only RF input is supported.
Suffix:
<ip>
.
1..n
Parameters:
<Source> RF
Radio Frequency ("RF INPUT" connector)
FIQ
I/Q data file
*RST: RF
Setting Basic Measurement Parameters