Manual operation: See " Radio Frequency State " on page 312
14.7.6.2 Setting up Probes
Modular probes can be connected to the RF input connector of the R&S FSV/A.
For details see Chapter 8.2.1.1, "Using Probes", on page 307.
[SENSe:]PROBe<pb>:ID:PARTnumber?.......................................................................... 925
[SENSe:]PROBe<pb>:ID:SRNumber?..............................................................................925
[SENSe:]PROBe<pb>:SETup:ATTRatio............................................................................926
[SENSe:]PROBe<pb>:SETup:CMOFfset.......................................................................... 926
[SENSe:]PROBe<pb>:SETup:DMOFfset.......................................................................... 926
[SENSe:]PROBe<pb>:SETup:MODE............................................................................... 927
[SENSe:]PROBe<pb>:SETup:NAME?..............................................................................927
[SENSe:]PROBe<pb>:SETup:NMOFfset.......................................................................... 928
[SENSe:]PROBe<pb>:SETup:PMODe............................................................................. 928
[SENSe:]PROBe<pb>:SETup:PMOFfset.......................................................................... 929
[SENSe:]PROBe<pb>:SETup:STATe?..............................................................................929
[SENSe:]PROBe<pb>:SETup:TYPE?...............................................................................929
[SENSe:]PROBe<pb>:ID:PARTnumber?
Queries the R&S part number of the probe.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<PartNumber>
Example: //Query part number
PROB3:ID:PART?
Usage: Query only
Manual operation: See "Part Number" on page 315
[SENSe:]PROBe<pb>:ID:SRNumber?
Queries the serial number of the probe.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<SerialNo>
Example: //Query serial number
PROB3:ID:SRN?