Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
929User Manual 1178.8520.02 ─ 01
Manual operation: See "Mode" on page 315
[SENSe:]PROBe<pb>:SETup:PMOFfset <PMOffset>
Sets the P-mode offset. The setting is only available if a modular probe in P-mode is
connected to the R&S FSV/A. The maximum voltage difference between the positive
and negative input terminals is 16 V.
If the probe is disconnected, the P-mode offset of the probe is reset to 0.0 V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see "MultiMode Function and Offset Compensation for Modular RF Probes"
on page 309.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<PMOffset> The voltage offset between the positive input terminal and
ground.
Default unit: V
Manual operation: See "Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
set / " on page 315
[SENSe:]PROBe<pb>:SETup:STATe?
Queries if the probe at the specified connector is active (detected) or not active (not
detected).
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Return values:
<State> DETected | NDETected
Example: //Query connector state
PROB3:SET:STAT?
Usage: Query only
[SENSe:]PROBe<pb>:SETup:TYPE?
Queries the type of the probe.
Setting Basic Measurement Parameters