Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
938User Manual 1178.8520.02 ─ 01
Example:
INIT:CONT OFF
Selects single sweep operation
CORR:METH TRAN
Selects a transmission measurement.
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Usage: Setting only
Manual operation: See "Calibrate Reflection Short" on page 342
See "Calibrate Reflection Open" on page 342
[SENSe:]CORRection:METHod
<Type>
This command selects the type of measurement to be performed with the external gen-
erator.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe] on page 933).
Parameters:
<Type> REFLection
Selects reflection measurements.
TRANsmission
Selects transmission measurements.
*RST: TRANsmission
Example:
CORR:METH TRAN
Sets the type of measurement to "transmission".
Manual operation: See "Calibrate Transmission" on page 342
See "Calibrate Reflection Short" on page 342
See "Calibrate Reflection Open" on page 342
[SENSe:]CORRection:RECall
Manual operation:
See " Recall " on page 342
[SENSe:]CORRection[:STATe] <State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with [SENSe:]CORRection:COLLect[:ACQuire] on page 937.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe] on page 933).
Parameters:
<State> ON | OFF | 0 | 1
Setting Basic Measurement Parameters