Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
968User Manual 1178.8520.02 ─ 01
Parameters:
<History> The maximum number of frames depends on the number of
sweep points.
Range: 781 to 20000
Increment: 1
*RST: 3000
Example:
CALC:SGR:SPEC 1500
Sets the history depth to 1500.
Manual operation: See " History Depth " on page 509
CALCulate<n>:SGRam:LAYout <State>
CALCulate<n>:SPECtrogram:LAYout <State>
This command selects the state and size of spectrograms.
The command is available for result displays that support spectrograms.
Suffix:
<n>
.
Window
Example:
CALC4:SPEC:LAY FULL
Shows the spectrogram in window 4. The corresponding trace
diagram is hidden.
Manual operation: See "State" on page 508
CALCulate<n>:SGRam[:STATe] <State>
CALCulate<n>:SPECtrogram[:STATe] <State>
This command turns the spectrogram on and off.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:SGR ON
Activates the Spectrogram result display.
CALCulate<n>:SGRam:THReedim[:STATe] <State>
CALCulate<n>:SPECtrogram:THReedim[:STATe] <State>
Activates or deactivates a 3-dimensional spectrogram for the selected result display.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 0 | 1