Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
969User Manual 1178.8520.02 ─ 01
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
CALC:SPEC:THR:STAT ON
Manual operation: See "3D Spectrogram State" on page 508
CALCulate<n>:SGRam:TRACe <Trace>
CALCulate<n>:SPECtrogram:TRACe <Trace>
Selects the spectrum diagram trace on which the spectrogram is based.
Suffix:
<n>
.
Window
Parameters:
<Trace> TRACe1 | TRACe2 | TRACe3 | TRACe4 | TRACe5 | TRACe6
Example:
CALC:SPEC:TRAC TRACe1
CALCulate<n>:SGRam:TSTamp:DATA? <Frames>
CALCulate<n>:SPECtrogram:TSTamp:DATA? <Frames>
This command queries the starting time of the frames.
The return values consist of four values for each frame. If the Spectrogram is empty,
the command returns '0,0,0,0'. The times are given as delta values, which simplifies
evaluating relative results; however, you can also calculate the absolute date and time
as displayed on the screen.
The frame results themselves are returned with TRAC:DATA? SGR
See TRACe<n>[:DATA] on page 975.
Query parameters:
<Frames> CURRent
Returns the starting time of the current frame.
ALL
Returns the starting time for all frames. The results are sorted in
descending order, beginning with the current frame.
Return values:
<Seconds> Number of seconds that have passed since 01.01.1970 till the
frame start
<Nanoseconds> Number of nanoseconds that have passed in addition to the
<Seconds> since 01.01.1970 till the frame start.
<Reserved> The third value is reserved for future uses.
<Reserved> The fourth value is reserved for future uses.