Remote Control Commands
R&S
®
SMB100A
391Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:PULM:DOUBle:DELay <Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay> float
Range: 10 ns to 100 s
Increment: 5 ns
*RST: 3 us
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
Options: R&S SMB-K23 (Pulse Generator)
Manual operation: See "Double Pulse Delay - Pulse Generator" on page 228
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Enables/disables double pulse generation. The two pulses are generated in one pulse
period.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
Options: R&S SMB-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:DOUBle:WIDTh <Width>
Sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width> float
Range: 10 ns to 100 s
Increment: 10 ns
*RST: 3 us
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Options: R&S SMB-K23 (Pulse Generator)
Manual operation: See "Double Pulse Width - Pulse Generator" on page 228
[:SOURce<hw>]:PULM:MODE <Mode>
Sets the mode of the pulse generator.
SOURce Subsystem