Fernsteuerbefehle
R&S
®
RTB2000
474Bedienhandbuch 1333.1611.03 ─ 05
Suffix:
<p>
.
1..2
Selects the logic pod
Parameter:
<State> ON | OFF
*RST: OFF
DIGital<m>:DISPlay <State>
Enables and displays the indicated logic channel, or disables it.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameter:
<State> ON | OFF
*RST: OFF
DIGital<m>:TECHnology <ThresholdMode>
Selects the threshold voltage for various types of integrated circuits and applies it to
the channel group to which the indicated logic channel belongs.
Suffix:
<m>
.
0..15
Number of the logic channel
Parameter:
<ThresholdMode> TTL | ECL | CMOS | MANual
TTL: 1.4 V
ECL: -1.3 V
CMOS: 2.5 V
MANual: Set a user-defined threshold value with DIGital<m>:
THReshold
*RST: MAN
LOGic<p>:THReshold:UDLevel
LOGic<p>:THReshold
Selects the threshold voltage for the indicated pod.
Alternative command for DIGital<m>:TECHnology.
Suffix:
<p>
.
1..2
Selects the logic pod
Mixed-Signal-Option (Option R&S RTB-B1)

Table of Contents