EasyManua.ls Logo

SYSMEX XP Series - Analysis Parameters

SYSMEX XP Series
290 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sysmex XP-300 1-5
Introduction
Revised October 2012
Unit
• dL: deciliter (0.1 liter)
• fL: femtoliter (10
-15
liter)
•μL: microliter (10
-6
liter)
• pg: picogram (10
-12
gram)
Analysis parameters
This instrument provides results for the following parameters:
• WBC: Number of leukocytes (Analysis principle: DC detection
method)
WBC count in 1 μL of whole blood
• RBC: Number of erythrocytes (Analysis principle: DC
detection method)
RBC count in 1 μL of whole blood
• HGB: Hemoglobin (Analysis principle: Non-cyanide
hemoglobin analysis method)
Concentration (gram) of hemoglobin in 1 dL of whole blood
• HCT: Hematocrit value (Analysis principle: RBC pulse height
detection method)
Ratio (%) of whole RBC volume in whole blood
• MCV: Mean RBC volume
Mean RBC volume (fL) in whole blood. The volume is
calculated by HCT/RBC.
• MCH: Mean RBC hemoglobin
Mean hemoglobin concentration (pg) per RBC. The
concentration is calculated by HGB/RBC.
• MCHC: Mean RBC hemoglobin concentration
Mean hemoglobin concentration (g/dL). The concentration is
calculated by HGB/HCT.
• PLT: Platelet (Analysis principle: DC detection method)
Platelet count in 1 μL of whole blood
• LYM% (W-SCR): WBC-Small Cell Ratio
Ratio (%) of small cells (the lymphocytes) to whole WBC
• MXD% (W-MCR): WBC-Middle Cell Ratio
Ratio (%) of middle cells (the mixed population of EO, BASO
and MONO) to whole WBC
• NEUT% (W-LCR): WBC-Large Cell Ratio
Ratio (%) of large cells (the neutrophils) to whole WBC
• LYM# (W-SCC): WBC-Small Cell Count
Absolute count of WBC-small cells (the lymphocytes) in 1 μL
of whole blood
• MXD# (W-MCC): WBC-Middle Cell Count
Absolute count of WBC-middle cells (the mixed population of
EO, BASO and MONO) in 1 μL of whole blood

Table of Contents